RE-52A旋轉蒸發儀標準操作規程

1、目的:建立RE-52A旋轉蒸發儀的操作規程,確保正確使用。 2、范圍:適用于RE-52A旋轉蒸發儀的標準操作。 3、責任: 
3.1 QC室檢驗員負責按照本規程的方法操作。 3.2 QC室主任負責監督檢查本規程的實施。 4、內容: 
4.2操作 
4.2.1 接通電源:五芯插頭插入箱后插座,三芯插頭插入220V/50HZ插座。 4.2.2 調正主機頭高度:先松開控制箱背后T形螺釘,左手托住機頭,右手轉動升降手輪,達需要高度后,停止轉動,并將T形螺釘旋緊。 
4.2.3 調正主機頭角度:旋松十字螺釘,主機頭、玻璃件、蒸發瓶連成一體的傳動蒸發系統就能沿水平方向在0-45°范圍內任意傾斜,達到你需要的角度后,旋緊十字螺釘即可。 
4.2.4 按電機開關:綠燈亮,主機開始旋轉。轉動調速旋鈕,主機0-130轉/分范圍內調速。 
4.2.5 按電熱開關水浴鍋自動加熱,恒溫控制。不許無水干燒 4.2.6 工作結束,關閉開關,拔下三芯插頭,使儀器斷電。 4.3維護保養 
4.3.1玻璃件:有嚴重污垢時,可用稀鹽酸清洗 
4.3.2四氟接口:一般情況不用拆洗,如有嚴重污垢可拆洗。水浴鍋內清水應經常調換、清洗。 
4.3.3密封圈:經常檢查密封圈的磨損(腐蝕)程度,及時更換磨損的密封圈,以確保儀器的真空度 
4.4故障分析與排除