<var id="zftvf"></var>
<cite id="zftvf"><strike id="zftvf"></strike></cite><cite id="zftvf"></cite>
<cite id="zftvf"><video id="zftvf"><thead id="zftvf"></thead></video></cite>
<menuitem id="zftvf"></menuitem>
<var id="zftvf"></var>
<menuitem id="zftvf"></menuitem><var id="zftvf"></var>
<var id="zftvf"></var>
<var id="zftvf"></var>
<cite id="zftvf"><video id="zftvf"></video></cite>
<cite id="zftvf"></cite>
<cite id="zftvf"></cite><var id="zftvf"></var><cite id="zftvf"><video id="zftvf"></video></cite>
<var id="zftvf"><video id="zftvf"></video></var><cite id="zftvf"><strike id="zftvf"></strike></cite>
<cite id="zftvf"><video id="zftvf"></video></cite><thead id="zftvf"><video id="zftvf"><thead id="zftvf"></thead></video></thead>
<ins id="zftvf"><span id="zftvf"></span></ins><ins id="zftvf"><noframes id="zftvf"><menuitem id="zftvf"></menuitem>
<cite id="zftvf"><video id="zftvf"><thead id="zftvf"></thead></video></cite>
<var id="zftvf"><video id="zftvf"><thead id="zftvf"></thead></video></var>

儀器使用操作規程介紹—旋轉蒸發儀

注意事項 
? 減壓旋轉蒸發,需要與循環水式真空泵連接,所以開機時要特別注意檢查各組件連接的密閉性; 
? 旋轉蒸發瓶和接收瓶均是玻璃磨口,且均有掐口固定,故操作時勿過分用力。 開機前的準備工作 
? 將水浴鍋和真空泵內添加適量的蒸餾水; 
? 檢查旋轉蒸發儀冷凝管部分是否有污物,若有,清理后再使用; ? 檢查冷凝水膠管和真空管是否連接緊密,以免漏氣; ? 提前打開水浴鍋開關,進行預加熱; 
? 旋轉蒸發瓶內料液體積不得超過瓶體積的2/3。 操作流程 
1、 安裝旋轉蒸發瓶和接收瓶; 
2、 連接電源插座,調整水浴鍋到指定溫度; 3、 打開冷凝水開關; 
4、 降下旋轉架,逆時針旋轉90°鎖死,打開旋轉按鈕,旋轉到一
定轉速,在水浴鍋升溫的同時,加熱瓶內料液; 
5、 待升**指定溫度,關閉排氣閥,打開真空泵,開始計時旋蒸; 6、 待旋蒸結束,關閉旋轉開關,再關閉真空泵以及冷凝水閥,然
后緩慢打開排氣閥,再關閉水浴鍋; 
7、 待氣體排出后,順時針旋轉90°解鎖,升起旋轉架,先取下旋
轉蒸發瓶,再取下接收瓶,關閉電源

設備清洗及維護 
將旋轉蒸發瓶內裝入2/3無水乙醇,水浴溫度50℃,重復上述步驟進行清洗,時間0.5 h。如遇到旋轉軸玻璃管兩端有難以清洗污物,則需找儀器負責人進行拆機清洗維護。 出現緊急情況時的應急措施 ? 旋蒸瓶噴液 
微調排氣閥,待壓力和溫度穩定后再關緊排氣閥。如果噴液過多,則須找儀器負責人對儀器進行拆機清洗; ? 旋蒸瓶擰不下來 
1、先檢查排氣閥是否開啟,若無,則先打開排氣閥進行排氣后再次嘗試。如果還是打不開,則進行如下操作; 
2、一定不要硬擰,注意連接旋蒸瓶的掐口,該掐口上部螺圈與旋轉軸固件以螺紋連接,順時針扭動螺圈使其向下頂碰旋蒸瓶,慢慢下壓,使旋蒸瓶被緩慢壓出磨口段; ? 真空泵達不到要求的壓力 
直接找儀器負責人。